Accreditation marks on a peptide test report: A2LA, ANAB and the ILAC/IAF merger, assessed

A logo on a certificate of analysis is not a purity grade. Assessed: what an ISO/IEC 17025 accreditation attests, why scope is the whole of it, and why the mutual-recognition line printed on reports is mid-rename.

By Imogen Slater Assessments Documentation Regulatory Terminology 1,132 words 6 min read

A research-supply certificate of analysis now tends to arrive with a badge on it. Somewhere near the purity figure sits an accreditation mark — most often, on a US-issued report, A2LA or ANAB — and buyers read it the way they read the percentage: as a grade. It is not a grade. It is a statement about a laboratory’s competence within a written scope, issued by a third party, and it says nothing at all about the number printed beside it. This desk assesses the mark as documentation.

What the mark attests

An accreditation body assesses a laboratory against ISO/IEC 17025, the international standard for testing and calibration laboratories, and grants accreditation for named activities. The word doing the work is scope. Accreditation is not held by an organisation in the abstract; it is held for particular tests, by particular methods, in particular disciplines.

A2LA’s own site makes the structure legible. Alongside its ISO/IEC 17025 testing and calibration laboratory accreditation it lists a long series of separately named programmes — Chemical Testing, Biological Testing, the AOAC Laboratory Accreditation Program, Cannabis Testing, Laboratory Accreditation for Analyses of Foods, the FDA ASCA pilots, WADA Testing, Calibration — plus further accreditation streams under ISO/IEC 17020 for inspection bodies, ISO/IEC 17065 for product certification, ISO/IEC 17043 for proficiency testing providers, ISO 17034 for reference material producers, ISO 15189 and CLIA for clinical laboratories, and ISO 20387 for biobanking. That is a catalogue of scopes, not a seal of general quality.

No research-peptide programme appears on that published list. The desk states that as what it is — an absence from A2LA’s own published programme catalogue, not a claim that no laboratory anywhere may hold a scope covering peptide analysis. A chemical-testing scope written around chromatographic identity and purity work can plainly reach a peptide. The point is that a reader cannot infer it from a logo; the scope document either names the method or it does not.

The wording is mid-rename

The transition is the live part. ILAC and IAF — the international cooperations behind, respectively, laboratory accreditation and management-system certification — have established a merged body, and ILAC’s site now heads itself Global Accreditation Cooperation Incorporated and announces its official launch, alongside a short history of the decision and the election of an incoming chair. This desk did not read a launch date and does not print one.

What matters for paperwork is that the same site still runs “ILAC MRA and Signatories” in its navigation and still offers signatory search under the ILAC MRA heading. So both forms are correct at once for now, and a report citing the ILAC MRA is not thereby stale, nor is one citing the new body thereby wrong. Anyone building a checklist that pattern-matches on the phrase will get a false negative for the duration.

Verifying one, if you are shown one

Three fields, which a reprinted badge rarely carries in full: the accreditation body, the certificate number, and the laboratory’s legal name. Take those to the body’s own organisation search — A2LA’s site carries a Search Organizations entry point. For ANAB the same check starts on ANAB’s own site, whose ISO/IEC 17025 testing-laboratory page refused this desk’s automated fetch — a browser rather than a script. Then read the scope document, not the certificate, and confirm the method used on your report appears in it.

The failure mode is not forgery. It is a genuine mark, genuinely held, for work that has nothing to do with the assay in front of you — and a supplier reprinting a laboratory’s letterhead as though the accreditation attached to the vial. Which laboratory ran the work is its own question, and one that moves: a contract testing house can change hands mid-year without a line of the report format changing.

Pros and cons

Verdict

The mark earns its place only once the certificate number and the scope have been checked. Until then it is a picture. Everything above concerns the paperwork behind materials supplied for laboratory research use only; nothing here is about what any compound does, and no compound discussed is intended for human consumption.